尊龙凯时[中国]官方网站

Product catalog
Contact Us
Online Message
Detail

High flow humidification treatment instrument

The high-flow humidification treatment device is a treatment device that provides continuous positive pressure and heating and humidification. Air and oxygen are filtered by filter cotton and inhaled by the fan. After passing through the mixing chamber, the air-oxygen mixture is formed, and then output to the humidification. In the tank, the surface of the hot water is heated and humidified to reduce the stimulation of mechanical ventilation on the cardiopulmonary system, keep the alveoli moist, facilitate sputum suction and prevent airway obstruction, and control the heating of the heating plate by setting the target gas temperature value. This temperature has real-time monitor. After heating and humidification, the heating wire breathing circuit is used to output the patient through the nasal tube after the second temperature rise.

Applications:

The high-flow respiratory humidification treatment instrument is used for patients with spontaneous breathing, and effective treatment is carried out by providing a certain flow, warming and humidifying breathing gas. These patients include patients undergoing humidification therapy, oxygen therapy, tracheal intubation, and tracheotomy. Cannot be used for life support equipment.

Features

● Three levels of temperature control, overheating (dry burning) protection, safe and effective.

● The measured values of flow, temperature, oxygen concentration and working status are displayed on the same screen at the same time.

● Flow rate automatic compensation function.

● The machine is equipped with high-density filter cotton to avoid cross-infection.

● The oxygen concentration can be adjusted in the whole range, no need to add an external oxygen flow meter.

● Integrated ultrasonic real-time oxygen concentration monitoring, oxygen-free battery consumption.

● Easy disinfection, only need to use common water-soluble disinfectant to moisturize, soft cloth to clean the surface of the machine and the surface of the parts that may be contaminated.   

technical parameter

Output gas temperature: 31℃~37℃ tolerance ±2%

                     1st gear 31℃ 2nd grade 34℃ 3rd grade 37℃

Flow adjustment range: LPM~80 LPM tolerance ±8%

Oxygen concentration adjustment monitoring range: 21%~100%, tolerance ±5% (±3% FS)

Heating plate thermal cut: 115±5℃

Automatic water-adding humidification tank: automatic water-adding, variable volume ≤280mL, compliance ≤0.4mL/cmH2O

Audible alarm: ≥55dB (A)

Preheating time≤30min, reaching 31℃≤10min

Working environment temperature: 18℃~26℃, relative humidity≤80%

Power supply AC90V~260V 47Hz~63Hz

Power 150W±15%

Accessories

                                adaptors


                Heated chambers


                      Patient interface


Website links
>1688 alibaba
> alibaba(Head Office
> alibabaShenzhen Branch

> 3618med

>pigeon-medical

>Made in China

>MedicalExpo en
>MedicalExpo cn
友情链接: